Classic DU/D

  Description  

 

Classic DU15 / Classic DU20 / Classic DU25

 

All-in one system

Features

ASTM Type I & Type II water/Clinical laboratory reagent water (CLSI). 

On-line real time TOC monitoring

Dual wavelength UV-lamp 254nm & 185nm

U-cloud platform for remote monitoring (Wifi module)

DI resin module

USB access port for data logging

 

 

Classic D15 / Classic D20 / Classic D25

 

ASTM Type II high purified water system Clinical Laboratory Reagent Water (CLSI)

Features

Integrated water dispenser arm 

U-cloud platform for remote monitoring (Wifi module)

DI resin module

USB access port for data logging

Specifications

 

  All-in-one system Type II water
Main Configurations DU15 DU20 DU25 D15 D20 D25
Pre-treatment module YES YES
Main host YES YES
Pre-guard cartridge A YES YES
Pre-guard cartridge B YES YES
Two-stage Reverse Osmosis YES YES
DI Resin YES YES
60L Water tank, stepless water level sensor YES YES
254nm UV light of water tank YES YES
Inlet air filter of water tank YES YES
Water leakage protection sensor YES YES
TOC Monitoring YES NO
Dual Wavelength (254nm & 185nm) UV Lamp YES NO
Ultra Protection Cartridge A YES NO
One Integrated Water Dispenser Arm YES YES
* Optional ultra purification cartridge B special for semiconductor industry that has higher deionization requirements

 

 

  All-in-one system Type II water
Model DU15 DU20 DU25 D15 D20 D25
Feed Water Requirement
Source Portable Tap Water
Conductivity < 2000µS/cm
TOC < 1ppm
Hardness* < 450ppm as CaCO₃
Pressure 0.1 ~ 0.4Mpa ( 7-72psi )
Temperature 5 ~ 45℃
PH 4 - 10
Type II High Purified Water
Resistivity at 25℃** > 8MΩ.cm; tipically 10-15MΩ.cm
Conductivity at 25℃** Tipically 0.1µS/cm
TOC < 30ppb (µg/L)
Particulates with size > 0.22µm*** < 1pc/µL
Endotoxin (Pyrogens)*** < 0.001 EU/mL
Bacteria*** < 0.01 cfu/mL (<10cfu/L)
Rnase*** < 1pg/mL
Dnase*** < 5pg/mL
Proteases*** < 0.15µg/mL
RO Rejection ≥ 99%
DI ion Rejection ≥ 99%
Production flow rate 15L/H 20L/H 25L/H 15L/H 20L/H 25L/H
Manual Control Water Flow Rate Maximum 2L/min, stepless control of flow rate
Quantitative Water Dispense Range 0.01L ~ 60L
Type I Ultra Purified Water
Resistivity at 25℃** 18MΩ.cm  
Conductivity at 25℃ 0.055µS/cm  
TOC ≤2ppb (µg/L)  
Particulates with size > 0.22µm*** No particles  
Endotoxin (Pyrogens) < 0.001 EU/mL N/A
Bacteria < 0.01 cfu/mL (<10cfu/L)  
Rnase/Dnase Free  
Proteases < 0.15µg/mL  
Manual Control Water Flow Rate Maximum 2L/min, stepless control of flow rate  
Quantitative Water Dispense Range 0.01L ~ 60L  
Electrical Requirement  
Electrical Voltage 110V/220V ±10%
Electrical Frequency 50Hz/60Hz
Power 135W
Size Information  
Pre-treatment Module 8.5kg
Main Host with
Dispenser Arm
28.4kg 29.1kg 29.6kg 26.5kg 27.7kg 27.7kg
30L Water Tank 9kg
60L Water Tank 10.66kg
100L Water Tank 12.2kg
External Dimension (WxDxH)
Pre-treatment Module 270 x 280 x 566 (mm)
Main host 328 x 540 x 600 (mm)
Integrated Dispenser Arm 42 x 85.5 x 231.5 (mm)
30L Water Tank 390 x 390 x 690 (mm)
60L Water Tank 390 x 390 x 930 (mm)
100L Water Tank 390 x 390 x 1235 (mm)
Packing Dimension  
Gross Weight
Pre-treatment Module 9.8kg
Main Host with
Dispenser Arm
39.2kg 39.9kg 40.4kg 38.5kg 39.2kg 39.7kg
30L Water Tank 12.4kg
60L Water Tank 14kg
100L Water Tank 15.6
Packing Dimension (WxDxH)
Pre-treatment Module 350 x 360 x 670 (mm)
Main Host with
Dispenser Arm
510 x 670 x 800 (mm)
30L Water Tank 520 x 520 x 780 (mm)
60L Water Tank 520 x 520 x 1020 (mm)
100L Water Tank 520 x 520 x 1325 (mm)
* If the tap water contains a high rate of Calcium ions and Magnesium ion which do not meet the requirement, please use salty box to remove part of calcium ions and magnesium ion in advance.
** Resistivity typically 10-15MΩ.cm at 25℃, Conductivity typically 0.1µS/cm, at 25℃.
*** Feed water quality should meet above requirements and purified water through the remote water dispenser with end 0.22µm filter.
Brochure

 